Disassemble the Radiation Shield — Temperature/Humidity
Station
Note: We recommend using a workbench or table to perform the following procedures.
1. Place the Temperature/
Humidity Station on a
level surface.
2. Loosen the three
mounting screws that
hold the mounting
bracket and the
radiation shield
together.
3. Separate the radiation
shield plates to locate
the Temperature/
Humidity sensor in the
plating. See “Remove
the Previous
Temperature/Humidity
Sensor” below for
instructions on
removing the previous sensor.
Remove the Previous Temperature/Humidity Sensor
1. Locate the Temperature/Humidity sensor on the second to bottom plate in
the radiation shield plating stack.
2. Remove the three screws and flat washer securing the Temperature/
Humidity sensor to the radiation plating.
Mounting
Bracket
Radiation
Shield
Assembly
Lock Washers
Flat Washers
4" Screws
Cable Clamp
T
emperature/Humidity
Sensor